Kelvin connection

What is Kelvin Connection?

Kelvin connection, also known as a 4-wire connection, is a technique used to measure the resistance of a low-resistance conductor. It involves using two pairs of wires to make an electrical connection to the device under test. One pair of wires carries the current, while the other pair senses the voltage drop across the device. By eliminating the resistance of the connecting wires, the Kelvin connection provides more accurate measurements than traditional 2-wire connections.

The Kelvin connection is commonly used in applications where accurate resistance measurements are critical, such as in the calibration of electrical instruments or in the testing of electronic components. It is also used in applications where the resistance of the connecting wires cannot be ignored, such as in high-current applications or when working with low-resistance materials.

Benefits of Kelvin Connection

The Kelvin connection provides several benefits over traditional 2-wire connections. By eliminating the resistance of the connecting wires, it provides more accurate measurements of low-resistance conductors. It also minimizes the effects of temperature and thermal gradients on the measurement, resulting in more stable and reliable readings.

In addition, the Kelvin connection offers better noise immunity, as the voltage sensing wires are shielded from the current-carrying wires. This makes it ideal for use in noisy environments or when measuring low-level signals.

Kelvin Connection Example

A common example of the Kelvin connection is in the measurement of the resistance of a shunt resistor in a power supply circuit. In this application, the shunt resistor is used to measure the current flowing through the circuit. A 4-wire connection is used to connect the shunt resistor to the measuring instrument, with one pair of wires carrying the current and the other pair sensing the voltage drop across the resistor. By using a Kelvin connection, the resistance of the connecting wires is eliminated, resulting in more accurate current measurements.

How to Implement Kelvin Connection

To implement a Kelvin connection, two pairs of wires are used to connect the device under test to the measuring instrument. One pair of wires carries the current, while the other pair senses the voltage drop across the device. The voltage sensing wires are connected as close as possible to the device terminals, while the current-carrying wires can be located further away.

Care should be taken to ensure that the connection wires are of the appropriate gauge and material to minimize resistance and thermal effects. Shielded twisted-pair wires are often used for voltage sensing to provide better noise immunity. Finally, the measuring instrument should be set up to use a 4-wire measurement mode to take advantage of the Kelvin connection.

header - logo

The primary purpose of this project is to help the public to learn some exciting and important information about electricity and magnetism.

Privacy Policy

Our Website follows all legal requirements to protect your privacy. Visit our Privacy Policy page.

The Cookies Statement is part of our Privacy Policy.

Editorial note

The information contained on this website is for general information purposes only. This website does not use any proprietary data. Visit our Editorial note.

Copyright Notice

It’s simple:

1) You may use almost everything for non-commercial and educational use.

2) You may not distribute or commercially exploit the content, especially on another website.